Technical electrophysiological problem before the physiology...

RSS Feed

Would you like to save this topic, event, protocol or job so you can find it again easily?

Just click the "Save to My Lab Drawer" link and the item will be saved in the My Lab Drawer section of your bench space.

Available to members only. Please log in or register for your free account now.

Bruder Jakob
Germany

Send PM
See Mini bio

Status: Frog Egg
Frog Egg
Topic Started by Bruder Jakob
on 10/27/2008 7:50 AM   
Reply to this post Go to the top of the page

Hello,

I have just finished mounting a set-up and configuring the amplifier (Heka, EPC7) with the model cell. However, upon entering the patch pipette (1mm, borosilicate, filled with 2M NaCl) into the bath (filled with NaCl 0.9%) despite the voltage being at 0mV, I obtain a big current offset where the current signal is out of range and saturates (flatline), and cannot be brought back to 0 with the Vhold knob or the Vp offset knob.
Furthermore, looking at the test pulse of the amplifier window (of the Patchmaster software) in the set-up and on-cell configurations, the current is now at 0mV but the voltage is this time out of range and cannot be zeroed.
Therefore, the 2 signals cannot be zeroed at once, as should be the case when starting an experiment.

Usually, when offset problems occur, I know that it can be due to the microelectrode, or the intra-cellular solution, or the bath ground. However, this time these do not seem to be implicated since they have been changed many times.

Here is a list of all parameters we have changed and findings:

-Amplifier: changed with another EPC7 amplifier - signals still out of range
-Headstage: changed with another EPC7 headstage - signals still out of range
-AgCl microelectrode: tried to chloride it with Chlorox, pure sodium hypochlorit, HCl and current - signals still out of range
-Pipettes: changed the resistance high , low - signals still out of range
-Intracellular solution: started with NaCl 2M, then saline... signals still out of range
-Bath ground: platinum multi wire cable touching the bath solution and connected to the headstage by a gold pin connector, and copper single wire cable touching the bath solution and connected to the headstage by a silver pin connector. - signals still out of range
-Bath solution: started with saline, NaCl 2M, PBS. Deionized water was the only solution to bring signals approximately to 0mV and pA, which is normal as it does not conduct. This suggests to me that the system must be ok and that the problem lies in the bath solution.
-We also tried directly connecting a 5 Ohm resistance onto the BNC connector of the Heka headstage touching the bath solution instead of the holder and pipette. This permit us to zeroe both signals suggesting that the problem lies within the holder/pipette/intracellular solution/microelectrode system...contrary to the previous point.
-The set-up has been grounded and signals are clear with the model cell. There does not seem to be any excessive noise.

So what can be the problem???

Thank you for help


Last edited Oct 27, 2008, 10:42 AM by Bruder Jakob
Replies
frasermoss
United States

Send PM
See Mini bio

Status: Neuroscience, Electrophysiology and Imaging Moderator
Frog Laureate
Posted By frasermoss
on 11/3/2008 11:31 AM   
Reply to this post Go to the top of the page

Is your recording chamber completely clean?

We recently had a problem that we could not maintain a steady holding voltage and determined that it was because the chamber had not been cleaned properly and there were some crystals of one ionic salt or another that was redissolving in the bath and messing with the solution compositions.



As a Scientist Solutions member, you are able to register a positive vote for any topic which you believe is useful and relevant to our board or any reply which you believe is especially well worded and helpful.

By participating in the voting, you will be helping to identify the best topics & replies on the board.

You may vote once for any one post, and you may not vote for your own posts.

A post (topic or reply) will earn one "thumbs up" icon for every 10 votes received (up to 3 thumbs up), and the person who made the post will also earn two bonus points.

learn more about member points.



Click here to
Become a member & join our
community (It's easy & free)
Already a member? Please log in
User Name  
Password  
Forget Password?
Find out about our Rewards System
Scientists
Not finding the answer you need?

Post a new topic

You must be logged in to post. Log in above.
Not a member yet? Click here to register
(it's free)
Thank You to Our Sponsor