Award
 » Home » Analytical Chemistry » Spectrometry (Atomic, Electron, Mass) » Helium Leak
 
Solutions Search! The Customized Life Science Search Engine
Search Site
Search Suppliers
Search Internet
Search over 6000 life science websites specifically selected by our expert scientist moderators.

Other Topics
9/30/2008 10:19 AM
kindly give some points o ...
9/30/2008 09:45 AM
No Fragmentation after MS ...
9/23/2008 02:14 PM
Archon Autosampler Volati ...
9/14/2008 09:54 PM
SPECTROMETER
7/8/2008 01:16 PM
ICP-MS Recommendations fo ...
6/10/2008 10:39 AM
Troubleshooting low relat ...
5/22/2008 11:53 PM
Tandem mass spectra
3/21/2008 03:47 AM
CdS Determination Via ICP ...
3/18/2008 08:44 PM
Thermo Fisher Scientifics ...
2/27/2008 06:00 PM
Thermo Fisher Scientific ...
2/18/2008 07:22 PM
Aspectrics Appoints Scien ...
2/14/2008 07:48 PM
Thermo Scientific LTQ Orb ...
8/3/2007 08:20 AM
ICPs
7/27/2007 08:29 PM
Argon filter for ICP-OES
12/5/2006 07:30 AM
ICP MS interference
11/13/2006 03:13 AM
mesh used in the mass spe ...
9/15/2006 10:25 PM
GFAAS trouble
8/18/2006 01:13 AM
trace metals in oils usin ...
7/4/2006 10:29 PM
FAAS for Manganese
5/4/2006 02:43 PM
MS for API release testin ...
3/30/2006 06:10 PM
E BOOK
1/31/2006 04:14 AM
Cadmium on GFAAS
1/23/2006 03:08 PM
Boron on gfaas
11/3/2005 07:11 PM
measurement of biomarkers ...
7/27/2005 09:31 AM
Estimation of Mercury in ...
5/24/2005 09:41 AM
N-terminal sequencing/iTR ...
1/25/2005 05:28 PM
MS for amino acids
Subscribet to topic
bottom of page RSS Feed Topic Feed
 Helium Leak [View Printable]
tredd

Frog Egg

See
Similar
Scientists





Group: Member
Posts: 2
Joined: Jun 17, 2008







 Send a personal messsage to tredd Reply with a quote from this post Go to the top of the page

I am a GCMS analyst for an Environmental lab that tests water and soils for volatile organic compounds. I recently changed the block on the back of my archon autosampler and now i have a helium leak on my soil purge needle. However the soil purge needle is not in use and should not be getting any He through it, let alone leaking. I attempted to remove the needle and cap it, but that did not work. we are still seeing a leak somewhere in the unit. any ideas on how this may be fixed?
.........................

 Posted Jun 17, 2008, 17:11 PM Last edited Jun 17, 2008, 17:16 PM by tredd
Dr. Analytical

Frog Laureate

See
Similar
Scientists



View Blogs


Group: Moderators
Posts: 37
Joined: Jun 09, 2008







 Go to homepage of Dr. Analytical Send a personal messsage to Dr. Analytical Reply with a quote from this post Go to the top of the page

Could you give us a little more detail about the method you are using, and anything else that might have changed?

.........................
Dr. Analytical

Posted Jun 18, 2008, 0:20 AM
tredd

Frog Egg

See
Similar
Scientists





Group: Member
Posts: 2
Joined: Jun 17, 2008







 Send a personal messsage to tredd Reply with a quote from this post Go to the top of the page

i am using method 8260B, but i actually found the problem lastnight before going home. When i had replaced the block on the back of the instrument i must have missed an o-ring on one of the valves. so the problem is fixed now.
.........................

Posted Jun 18, 2008, 10:15 AM
Dr. Analytical

Frog Laureate

See
Similar
Scientists



View Blogs


Group: Moderators
Posts: 37
Joined: Jun 09, 2008







 Go to homepage of Dr. Analytical Send a personal messsage to Dr. Analytical Reply with a quote from this post Go to the top of the page

Glad to hear that you were able to solve the problem. Visit us again whenever you have questions.
.........................
Dr. Analytical

Posted Jun 18, 2008, 12:50 PM
bbeck

Frog Egg

See
Similar
Scientists





Group: Member
Posts: 2
Joined: Sep 23, 2008







 Send a personal messsage to bbeck Reply with a quote from this post Go to the top of the page

In a related query, I am having trouble with carryover on soils with my Archon. The water method does not carryover. I can up the number of rinses on the water method, but not for soils. Any ideas? Possible helium rinse leak, but all flows seem fine.
.........................