liquid junction potential [View Printable]
|
lazy
Group: Member Posts: 15 Joined: Apr 03, 2008
|
Although there are some nice tools to calculate the lilquid junction potential, I 'd like to know how to obtain the value on the oscilloscope. I think when I dip the electrode into the external solution, the current base line does not stay on zero line (0). I add the voltage to bring back the current line to the zero line. I think the voltage that I added is the liquid junction potential. Please let me know if my understanidng is OK or not. Thanks.
|
.........................
|
| Posted Apr 30, 2008, 8:00 AM |
|
|
|
frasermoss
Group: Admin Posts: 613 Joined: Feb 22, 2005
|
You need to dip your pipette loaded with internal solution into some more internal solution that you have put in your chamber.
Zero the baseline offset
lift out your pipette and replace the solution in the chamber with your extracellular solution and dip the pipette back in.
The readout should now be your LJP.
|
......................... "Opportunity is missed by most people because it is dressed in overalls and looks like work". Edison
|
| Posted Apr 30, 2008, 11:40 AM |
|
|
|
jaga_ch
Group: Member Posts: 1 Joined: May 25, 2008
|
hi, I was wondering which software do you use fo patch clamp? I work with Tida (Heka) and I am also trying to find out how to corect junction potential, however as far as I read and understood, the idea of adding the junction potential to the holding potential is right. :)
|
.........................
|
| Posted May 25, 2008, 13:28 PM |
|
|
|
lazy
Group: Member Posts: 15 Joined: Apr 03, 2008
|
I use Pulse software.
Calculated by Kenyon's method, I obtained appx +10 mV of LJP.
(1) I simply input +10 to the LJP field, and then I do patch as the standard way. I am not sure this is correct or not. Note that I do not change the bath solution.
(2) First, I understood that the pipette has this value when dipped into the bath solution. So, if I set the HP to -70 mV in whole cell without LJP cancellation, then the real HP must be -60 mV (-70+10=-60).
(3) But, when the whole cell is achieved, the LJP should be disappaired! (I think most people agree with this). Then, for what we have to cancel the LJP? The real HP can be -80mV with LJP cancellation, while I believe it -70mV???
(4) But the manual says how to cancel the LJP, meaning that it is required for the process to achieve the whole cell configuration.
(5) Currently, I am thinking that the reason why we have to cancel the LJP is because Auto Vo reset includes the LJP value, indicating that the offset line (I believed 0 mV) is not 0 mV, but +10 mV. I need to cancel this value with the way that the manual says. I need not care about LJP disapprearance in whole cell configuration.
But I am not sure what I am doing is good or not. How about your case, dear patch clampers?
|
.........................
|
| Posted May 26, 2008, 9:19 AM |
|
|
|
frasermoss
Group: Admin Posts: 613 Joined: Feb 22, 2005
|
For those of you have problems understanfing the LJP this is a must read. Correction for liquid junction potentials in patch clamp experiments. Methods Enzymol. 1992;207:123-31. Neher E. This chapter describes corrections that have to be applied to measured membrane potentials in patch clamp experiments. Some of them [Eqs. (1)-(3)] are required regardless of the nature of the reference electrode (in the Ringer's solution bath) whenever the pipette-filling solution is different from the bath solution. They represent the liquid junction potentials that are present at the pipette tip before patch formation. In addition, corrections have to be applied when the bath solution is being changed during a measurement (i.e., after seal formation). In that case the following rules apply. (1) The new solution should never get into contact with the bare silver/silver chloride wire of the reference electrode. This requirement is best met by using a salt bridge. (2) The "best" salt bridge is a 3 M KCl bridge with an abrupt KCl-bath fluid boundary at its tip (see above). This bridge does not require any additional potential corrections, but it may lead to KCl poisoning of the bath or become contaminated by solutions used previously. (3) Local solution changes (microperfusion by puffer pipette, U tool or sewer pipe arrangements) as well as recessed KCl bridges require additional corrections, which (together with the simple liquid junction potential correction) are approximately given by Eqs. (6)-(8). It should be stressed that all equations given here represent approximate corrections, since liquid junction potentials are thermodynamically ill-defined. This is particularly relevant for Eqs. (6) and (7) where the sum of two liquid junction potentials appears. PMID: 1528115 Pubmed link : http://www.ncbi.nlm.nih.gov/pubmed/1528115?dopt=abstractOrder it from your library today!
|
......................... "Opportunity is missed by most people because it is dressed in overalls and looks like work". Edison
|
| Posted May 26, 2008, 15:06 PM |
|
|
|
frasermoss
Group: Admin Posts: 613 Joined: Feb 22, 2005
|
Furthermore, if you use pClamp, it has a built in piece of software "JPCalc" that will give a best estimate of your LJP in your experiment if you give it all the parameters you will be using. For a reference see J Neurosci Methods (1994) 51: 107-16. Pubmed reference pageFor the stand alone JPCalc program go here http://web.med.unsw.edu.au/phbsoft/research_jpcalc.htm
|
......................... "Opportunity is missed by most people because it is dressed in overalls and looks like work". Edison
|
| Posted May 26, 2008, 15:15 PM |
|
|
|
lazy
Group: Member Posts: 15 Joined: Apr 03, 2008
|
I am considering 3 junction potentials:
(1) Junction potential between reference electrode and bath solution (2) Ag/AgCl wire in the pipette and pipette solution (3) liquid junction potential between pipette and bath solutions
In whole cell configuration, LJP will be disappeared.
Prior to patch process, I offset the potentials of (1) and (2), but not LJP (3), since it will disappear. For that, I give the value of the LJP calculated to the software.
|
.........................
|
| Posted May 27, 2008, 8:53 AM |
|
|
|
Konstanz
Group: Member Posts: 12 Joined: Aug 07, 2008
|
Hello, I have another question. Many people make correction for the LJP after the recordings. But in which cases it should be done before the recordings?
|
.........................
|
| Posted Aug 07, 2008, 6:53 AM |
|
|
|
frasermoss
Group: Admin Posts: 613 Joined: Feb 22, 2005
|
Personally I always apply the correction before. But I am 100% a voltage rather than current clamper.
Most amplifiers have separate controls for the offset and Vhold, so you can adjust the offset for zero current during the reference set up while Vhold is active and set to -Vlj (negative liquid junction potential).
this will anticipate the potential that will be seen by the patch after seal formation.
Vhold readings are then interpreted as -Vm for inside-out and Cell-attached patches or as +Vm for outside-out or whole-cell.
|
......................... "Opportunity is missed by most people because it is dressed in overalls and looks like work". Edison
|
| Posted Aug 07, 2008, 14:36 PM |
Last edited Aug 07, 2008, 14:41 PM by frasermoss |
|
|
|