Scientist Solutions: Life Science Discussions
 Refer a Friend    Link To Us    Bookmark Us       

      
 » Home » Analytical Chemistry » Chromatography » SHimadzu autosampler Question

Other Topics
11/20/2008 09:31 PM
Melamine analysis
11/3/2008 11:56 AM
IEX-HPLC for peptide mapp ...
10/29/2008 11:06 PM
Empty Capsule Testing by ...
10/16/2008 09:07 AM
Trace GC + Fisons AS800 a ...
10/6/2008 11:07 AM
formic acid HPLC
9/14/2008 04:03 PM
Queries
9/12/2008 10:10 PM
Formic Acid Quantitation
9/10/2008 06:20 AM
Formic acid quantitation
9/10/2008 06:26 AM
GC - Pump down
9/4/2008 09:11 AM
Base line Problems
8/28/2008 09:05 AM
chromatogram shape and in ...
8/19/2008 07:35 AM
detergent and lipids in F ...
8/14/2008 01:34 PM
Auto injector errors
8/10/2008 12:50 AM
HPLC for peptides
7/18/2008 10:06 AM
thiocyanate
7/16/2008 10:23 PM
A Good Site for Chemical ...
6/17/2008 01:01 PM
New Chromatography Softwa ...
5/29/2008 08:43 AM
GC FID detector with high ...
5/28/2008 06:49 PM
Dionex PAD-2
5/27/2008 09:36 AM
Does different HPLC syste ...
5/19/2008 06:36 AM
RP-HPLC chromatograph int ...
4/22/2008 02:07 PM
GC baseline problems
3/24/2008 08:31 PM
HPLC separation of colla ...
3/17/2008 02:10 PM
Ultra Performance Liquid ...
3/17/2008 02:12 PM
RPHPLC/ separation of sma ...
3/10/2008 08:27 PM
Thermo Fisher Scientific ...
2/17/2008 11:24 AM
Flow analysis and Flow sy ...
2/11/2008 06:19 PM
Thermo Fisher Scientific ...
12/6/2007 08:09 PM
Overly complex flow progr ...
7/31/2007 10:37 AM
Chiral separation of Dexi ...
Subscribet to topic
Add Reply  Add New Topic  Add New Poll
bottom of page RSS Feed 

Topic Feed

 

SHimadzu autosampler Question

 [View Printable]
efoconnor

Frog Egg

See
Similar
Scientists





Group: Member
Posts: 3
Joined: Mar 27, 2008







 Send a personal messsage to efoconnor Reply with a quote from this post Go to the top of the page

We have been producing blocked columns at an alarming rate. On investigation I noticed a pile of "chips' resembling graphite near the needle guide and on the needle spring assembly. Pursuing this further we noticed the "chip" on the plate seals and on the needle seal. We now belive these chips may be getting into the system. Has anyone else noted these 'chips"

Thanks
Ed

.........................

Posted Mar 27, 2008, 22:26 PM
DocJackman

Frog Egg

See
Similar
Scientists





Group: Member
Posts: 5
Joined: Feb 16, 2008







 Send a personal messsage to DocJackman Reply with a quote from this post Go to the top of the page

I have called this cokeing. It seems to occur when the injector is very hot and leaking air. It also seems to matter on volume and type of compounds being injected. 5ul or larger injections that are percent level PCB for instance seems to be able to produce significant carbon deposits. Also using ethylene glycol diacetate as your solvent can make it difficult to avoid. Frequent replacement of septa and good seals can help reduce the problem.

.........................

Posted Mar 30, 2008, 15:13 PM