Given you previous posts as well as this one I STRONGLY reccomend that read the following book
Patch Clamping
An Introductory Guide to Patch Clamp Electrophysiology
By Areles Molleman
Wiley Publishers ISBN 0-471-48685-X
Also read the Axon Guide
http://www.moleculardevices.com/pages/instruments/axon_guide.htmlFor cell attached patch you fill the pipette with the same solution that is bathing the cell (extracellular solution).
!!Remember !! in cell attached mode
1) Positive current = inward membrane current
2) membrane potential = Vrest - Vp
You are going to have to coat your pipettes with Sylgard #184 to keep the noise levels low.
Optimally you will need to use a headstage compatible with your amplifier that has a feedback resistance around 50GOhm. There is a probably a switch on your amplifier that allows you to change the feedback resistance (or capacitance) of your headstage circuit from "whole-cell" mode to "patch" mode.