Scientist Solutions: Life Science Discussions
 Refer a Friend    Link To Us    Bookmark Us       

      
 » Home » Analytical Chemistry » Spectrometry (Atomic, Electron, Mass) » CdS Determination Via ICP-OES

Other Topics
11/21/2008 02:37 PM
ABI 2000 and ThermoFinnig ...
9/30/2008 10:19 AM
kindly give some points o ...
9/30/2008 09:45 AM
No Fragmentation after MS ...
9/23/2008 02:14 PM
Archon Autosampler Volati ...
9/23/2008 02:35 PM
Helium Leak
9/14/2008 09:54 PM
SPECTROMETER
7/8/2008 01:16 PM
ICP-MS Recommendations fo ...
6/10/2008 10:39 AM
Troubleshooting low relat ...
5/22/2008 11:53 PM
Tandem mass spectra
3/18/2008 08:44 PM
Thermo Fisher Scientifics ...
2/27/2008 06:00 PM
Thermo Fisher Scientific ...
2/18/2008 07:22 PM
Aspectrics Appoints Scien ...
2/14/2008 07:48 PM
Thermo Scientific LTQ Orb ...
8/3/2007 08:20 AM
ICPs
7/27/2007 08:29 PM
Argon filter for ICP-OES
12/5/2006 07:30 AM
ICP MS interference
11/13/2006 03:13 AM
mesh used in the mass spe ...
9/15/2006 10:25 PM
GFAAS trouble
8/18/2006 01:13 AM
trace metals in oils usin ...
7/4/2006 10:29 PM
FAAS for Manganese
5/4/2006 02:43 PM
MS for API release testin ...
3/30/2006 06:10 PM
E BOOK
1/31/2006 04:14 AM
Cadmium on GFAAS
1/23/2006 03:08 PM
Boron on gfaas
11/3/2005 07:11 PM
measurement of biomarkers ...
7/27/2005 09:31 AM
Estimation of Mercury in ...
5/24/2005 09:41 AM
N-terminal sequencing/iTR ...
1/25/2005 05:28 PM
MS for amino acids
Subscribet to topic
Add Reply  Add New Topic  Add New Poll
bottom of page RSS Feed 

Topic Feed

 

CdS Determination Via ICP-OES

 [View Printable]
ShivaTecAnalyst

Frog Egg

See
Similar
Scientists





Group: Member
Posts: 1
Joined: Jul 19, 2007







 Send a personal messsage to ShivaTecAnalyst Reply with a quote from this post Go to the top of the page

Does anyone know the best digestion method for CdS for the determination of Sulfur? CdS digests readily in HCl however sulfur is lost during the formation of H2S. Any ideas would be appreciated.

.........................

Posted Jul 19, 2007, 20:29 PM
DocJackman

Frog Egg

See
Similar
Scientists





Group: Member
Posts: 5
Joined: Feb 16, 2008







 Send a personal messsage to DocJackman Reply with a quote from this post Go to the top of the page

If your source is cadmium sulfide; why not digest the sample in HCl and analyze for cadmium?

An alternative would be to digest the cadmium sulfide in a flask that is swept with a nitrogen or argon stream into a solution of standard zinc sulfate. After filtering and washing the precipitate, measure the final volume and analyze for residual zinc. The moles of zinc removed from the zinc sulfate should equal the number of moles of cadmium left in the original flask. This will also be the number of moles of sulfur released from the cadmiium.

You can always do the iodometric titration method which amounts to adding a known amount of standard iodine solution before acidifying the sample. The iodine will oxidize the sulfide quantitatively (you will need to prevent having too much oxygen from the air from accessing the sample). You then titrate the remaining iodine with standard thiosulfate until the solution has a light straw color. Add some starch indicator and continue to titrate until a colorless solution persists for ten seconds or more. With the proper calculations; you will have directly measured the sulfide sulfur.

.........................

Posted Mar 21, 2008, 3:16 AM