skgndri
India

Send PM
See Mini bio

Status: Tadpole
Tadpole
Topic Started by skgndri
on 5/28/2013 2:55 AM   
Reply to this post Go to the top of the page

We are observing a very high recovery of selenium in tap water . Kindly help.



Attached file: Se in WATER UM data March 2013.rar (3 downloads, 11KB)

Replies
Jruckle
United States

Send PM
See Mini bio

Status: Tadpole
Tadpole
Posted By Jruckle
on 5/29/2013 9:24 AM   
Reply to this post Go to the top of the page

I can not open your file.  Can you give more information about your problem.



skgndri
India

Send PM
See Mini bio

Status: Tadpole
Tadpole
Posted By skgndri
on 5/30/2013 1:55 AM   
Reply to this post Go to the top of the page

0.045 ml of 1 ppm selenium stock was spiked in 10 ml tap water and samples were analyzed on ICP MS ELAN 9000. Instrument was calibrated with 5 and 50 ppb standards. All samples and standards were having 0.2 ml of suprapur HNO3. Yettrium was added as internal standard. THe recovery found was 172 % in tap water samples. However, it was good in distilled water. Kindly suggest. Samples were not digested.



Jruckle
United States

Send PM
See Mini bio

Status: Tadpole
Tadpole
Posted By Jruckle
on 5/30/2013 2:29 AM   
Reply to this post Go to the top of the page

What mass are you using for selenium.  Kr in the argon can interfer with seleneum.  I have in the past had bad argon and if I tracked the Kr 83 mass it would start out with low counts and then after about an hour the counts would shoot up about ten fold.   This whould cause my Se to also shoot up.  Hope this helps.



skgndri
India

Send PM
See Mini bio

Status: Tadpole
Tadpole
Posted By skgndri
on 5/31/2013 0:18 AM   
Reply to this post Go to the top of the page

We are checking mass 82 with a correction equation given for krypton ( APHA 21st edition) C82-(1.008696*C83)



Jruckle
United States

Send PM
See Mini bio

Status: Tadpole
Tadpole
Posted By Jruckle
on 5/31/2013 7:41 AM   
Reply to this post Go to the top of the page

What else are you seeing in the samples?  You could be getting a Zinc oxide interference if zinc is high?  Sorry I cant think of any thing else.



Jruckle
United States

Send PM
See Mini bio

Status: Tadpole
Tadpole
Posted By Jruckle
on 5/31/2013 7:43 AM   
Reply to this post Go to the top of the page

What else are you seeing in the samples?  You could be getting a Zinc oxide interference if zinc is high?  Sorry I cant think of any thing else.



Sven1982
Belgium

Send PM
See Mini bio

Status: Froglet
Froglet
Posted By Sven1982
on 5/31/2013 10:30 AM   
Reply to this post Go to the top of the page

Se-82 is prone to interferences from ArCa (40+42 and 38+44)



As a Scientist Solutions member, you are able to register a positive vote for any topic which you believe is useful and relevant to our board or any reply which you believe is especially well worded and helpful.

By participating in the voting, you will be helping to identify the best topics & replies on the board.

You may vote once for any one post, and you may not vote for your own posts.

A post (topic or reply) will earn one "thumbs up" icon for every 10 votes received (up to 3 thumbs up), and the person who made the post will also earn two bonus points.

learn more about member points.