Scientist Solutions: International Life Science Community By Scientists For Scientists
    
Home » Forums » DNA (PCR, Real Time PCR) » Quantitative PCR (qPCR) from DNA Template » Problem in repeatability of Real time PCR results

Thanks to our sponsors who make this site possible

Problem in repeatability of Real time PCR results

RSS Feed

Would you like to save this topic, event, protocol or job so you can find it again easily?

Just click the "Save to My Lab Drawer" link and the item will be saved in the My Lab Drawer section of your bench space.

Available to members only. Please log in or register for your free account now.

phani9999
India

Send PM
See Mini bio

Status: Tadpole
Tadpole
Topic Started by phani9999
on 1/25/2012 23:33 PM   
Reply to this post Go to the top of the page

Dear Friends,

We are working on HBV quantification by using Rotor gene with Artus HBV PCR kit, the problem here is we are trying to work in 1/4 of the recommended volume and it is not showing similar results with the full volume results.

We have taken 7.5 microliters of master mix and 5 microliter of Extracted DNA and the recommended volume is 30 microliter of master mix and 20 microliter of extracted DNA

The results shows as follows.

For 12.5 microliter volume reaction copies are 89180253 Iu/ml.

For 50 microliter reaction copies are 166873792 Iu/ml.

What our idea is if you add 4 volumes greated amout of DNA in the 50 microliter reaction we are expecting 12.5 reaction copies * 4..But as per our results it is showing only less thandouble amount of copies

Please let us know the problem and solution.

Phani.


Replies
Ivan
United States

Send PM
See Mini bio

Status: DNA Moderator
Frog Laureate
Posted By Ivan
on 1/26/2012 10:48 AM   
Reply to this post Go to the top of the page

Hi Phani,

Getting linear results when you use one fourth of the recommended volume of a qPCR reaction is not common. In other words using lower volumes works, but it will never give you the robust signal that a 50 ul reaction does. And to put it in yet another way, you cannot really compare the signals obtained in a 50 ul reaction that obtained from a 12.5 ul reaction, even if both came from the same master mix

I would recommend running a standard curve at both reaction volumes and determine if both reaction volumes give you the same slope and y-intercept. Most likely they will not. In that case, you should use the reaction volume that gives you the best results, which will also be the reaction volume with the most accurate quantification

Best

Ivan Delgado Orlic Carlsbad, CA



phani9999
India

Send PM
See Mini bio

Status: Tadpole
Tadpole
Posted By phani9999
on 2/2/2012 2:09 AM   
Reply to this post Go to the top of the page

Dear Ivan,

Thank you.



As a Scientist Solutions member, you are able to register a positive vote for any topic which you believe is useful and relevant to our board or any reply which you believe is especially well worded and helpful.

By participating in the voting, you will be helping to identify the best topics & replies on the board.

You may vote once for any one post, and you may not vote for your own posts.

A post (topic or reply) will earn one "thumbs up" icon for every 10 votes received (up to 3 thumbs up), and the person who made the post will also earn two bonus points.