Topic Started by Tshee15
on 1/18/2012 10:57 AM   
Reply to this post Go to the top of the page

Is it normal for the DAC value to be:
Be 6.8
Co 8.5
In 8.5

and still pass the SmartTune Autolens?  I was assuming that the DAC values had to go up linearly.  Is this something that would affect Mn not passing a calibration curve????

Thanks in advance.


Replies
Jruckle
United States

Send PM
See Mini bio

Status: Frog Egg
Frog Egg
Posted By Jruckle
on 1/19/2012 5:53 AM   
Reply to this post Go to the top of the page

Sounds Like an Oxide problem.  try manualy lowering your Neb flow, than redo your Autolens.



Tshee15
United States

Send PM
See Mini bio

Status: Frog Egg
Frog Egg
Posted By Tshee15
on 1/19/2012 6:35 AM   
Reply to this post Go to the top of the page

Thank you for the reply.

What is the lowest number that the oxides could be?  In the manual it states you want to get as close as possible to 0.3 without going over.  I am sure that if you go too low than that would affect things as well.



Jruckle
United States

Send PM
See Mini bio

Status: Frog Egg
Frog Egg
Posted By Jruckle
on 1/19/2012 8:11 AM   
Reply to this post Go to the top of the page

the lower the better.  That said as you lower the Neb flow you will see the background go up. I try to keep that below 30 counts at mass 220



Tshee15
United States

Send PM
See Mini bio

Status: Frog Egg
Frog Egg
Posted By Tshee15
on 1/21/2012 5:27 AM   
Reply to this post Go to the top of the page

I lowered the neb flow to 0.84 and the Autolens is failing
Be 7.25  max intesity 1549
Co 10 m.i. 118037
In 10 m.i. 232514

All with 29 points.

Is it ok to continue on with the daily performance or does the autolens need to pass in order to proceed.

Thank you for all your help on this.  I am new to this instrument and do not have much guidance.



Jruckle
United States

Send PM
See Mini bio

Status: Frog Egg
Frog Egg
Posted By Jruckle
on 1/23/2012 3:42 AM   
Reply to this post Go to the top of the page

the lower your neb flow the lower your intensitys.  If you need to lower your neb flow that low to get a good oxid ratio you may have a leak some were in the system.  I would clean your sample intro system.  Nebulizer all the way to the lens.  Change all your o- ring. then redo your optimization.

hope this helps.



As a Scientist Solutions member, you are able to register a positive vote for any topic which you believe is useful and relevant to our board or any reply which you believe is especially well worded and helpful.

By participating in the voting, you will be helping to identify the best topics & replies on the board.

You may vote once for any one post, and you may not vote for your own posts.

A post (topic or reply) will earn one "thumbs up" icon for every 10 votes received (up to 3 thumbs up), and the person who made the post will also earn two bonus points.

learn more about member points.



Click here to
Become a member & join our
community (It's easy & free)
Already a member? Please log in
User Name  
Password  
Forget Password?
Find out about our Rewards System
Scientists
Not finding the answer you need?

Post a new topic

You must be logged in to post. Log in above.
Not a member yet? Click here to register
(it's free)
Thank You to Our Sponsor

The Latest Events