Award
 » Home » Analytical Chemistry » Spectrometry (Atomic, Electron, Mass) » GFAAS trouble
 
Solutions Search! The Customized Life Science Search Engine
Search Site
Search Suppliers
Search Internet
Search over 6000 life science websites specifically selected by our expert scientist moderators.

Other Topics
9/30/2008 10:19 AM
kindly give some points o ...
9/30/2008 09:45 AM
No Fragmentation after MS ...
9/23/2008 02:14 PM
Archon Autosampler Volati ...
9/23/2008 02:35 PM
Helium Leak
9/14/2008 09:54 PM
SPECTROMETER
7/8/2008 01:16 PM
ICP-MS Recommendations fo ...
6/10/2008 10:39 AM
Troubleshooting low relat ...
5/22/2008 11:53 PM
Tandem mass spectra
3/21/2008 03:47 AM
CdS Determination Via ICP ...
3/18/2008 08:44 PM
Thermo Fisher Scientifics ...
2/27/2008 06:00 PM
Thermo Fisher Scientific ...
2/18/2008 07:22 PM
Aspectrics Appoints Scien ...
2/14/2008 07:48 PM
Thermo Scientific LTQ Orb ...
8/3/2007 08:20 AM
ICPs
7/27/2007 08:29 PM
Argon filter for ICP-OES
12/5/2006 07:30 AM
ICP MS interference
11/13/2006 03:13 AM
mesh used in the mass spe ...
8/18/2006 01:13 AM
trace metals in oils usin ...
7/4/2006 10:29 PM
FAAS for Manganese
5/4/2006 02:43 PM
MS for API release testin ...
3/30/2006 06:10 PM
E BOOK
1/31/2006 04:14 AM
Cadmium on GFAAS
1/23/2006 03:08 PM
Boron on gfaas
11/3/2005 07:11 PM
measurement of biomarkers ...
7/27/2005 09:31 AM
Estimation of Mercury in ...
5/24/2005 09:41 AM
N-terminal sequencing/iTR ...
1/25/2005 05:28 PM
MS for amino acids
Subscribet to topic
bottom of page RSS Feed Topic Feed
 GFAAS trouble [View Printable]
James Ver.

Frog Egg

See
Similar
Scientists





Group: Member
Posts: 2
Joined: Jul 14, 2005







 Send a personal messsage to James Ver. Reply with a quote from this post Go to the top of the page

The graphite furnace atomic absorption unit I'm working with is acting up, and I was wondering if anyone here could help me try to fix it. The instrament is a Perkin Elmer Zeeman 5100, but I'm not sure how old it is.
The problem is that when I try to run a sample the instrament says that the tube burns out right before the vaporization stage. The tube will go through a conditioning run both before and after it's said to have burned out, so it can't be the tube. (if it's helpfull, the conditioning run goes up to 2650 degrees C, while the vaporization stage goes up to 2600 degrees.) The tube has been replaced, the contact cylenders have been raplaced, and the circut board has been fixed (there were a few minor problems with it; the machine's fairly old), but nothing seems to help. The only other thing anyone I've spoken with can think of is to take the unit apart and look at the wires themselves, and no one wants to do that because of the work it'd take to re-calibrate the unit.
Again, any help you could offer would be appreciated. Thank you.
.........................

 Posted Jul 14, 2005, 14:40 PM
Sandy

Frog Laureate

See
Similar
Scientists





Group: Member
Posts: 117
Joined: Nov 23, 2004







 Send a personal messsage to Sandy Reply with a quote from this post Go to the top of the page

James Ver. said:
The graphite furnace atomic absorption unit I'm working with is acting up, and I was wondering if anyone here could help me try to fix it. The instrament is a Perkin Elmer Zeeman 5100, but I'm not sure how old it is.
The problem is that when I try to run a sample the instrament says that the tube burns out right before the vaporization stage. The tube will go through a conditioning run both before and after it's said to have burned out, so it can't be the tube. (if it's helpfull, the conditioning run goes up to 2650 degrees C, while the vaporization stage goes up to 2600 degrees.) The tube has been replaced, the contact cylenders have been raplaced, and the circut board has been fixed (there were a few minor problems with it; the machine's fairly old), but nothing seems to help. The only other thing anyone I've spoken with can think of is to take the unit apart and look at the wires themselves, and no one wants to do that because of the work it'd take to re-calibrate the unit.
Again, any help you could offer would be appreciated. Thank you.


I think you should contact the manufacturing company: Perkin Elmer, click on the following URL:

http://las.perkinelmer.com/Contact+Us/default.htm?Region=North%20America
.........................

Posted Jul 14, 2005, 15:37 PM
James Ver.

Frog Egg

See
Similar
Scientists





Group: Member
Posts: 2
Joined: Jul 14, 2005







 Send a personal messsage to James Ver. Reply with a quote from this post Go to the top of the page

One of the professors here is going to do that. We just hit a point where we didn't know what else to do, and we figured someone else may have run into the same problem.
.........................

Posted Jul 14, 2005, 15:39 PM
pipn

Frog Egg

See
Similar
Scientists





Group: Member
Posts: 2
Joined: Sep 15, 2006







 Send a personal messsage to pipn Reply with a quote from this post Go to the top of the page

it's the wires I've had this happen before if your handy with tools you can fix it I had to dismantle the magnet surounding the furnace and the replace then reroute the faulty wire. seems the opening and closing of the door over time caused it to fail


If too much at stake if you go into the machine call p/e will cost $$$$
.........................

Posted Sep 15, 2006, 22:44 PM