<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description>Loud Noisy Roughing Pump Communication Error Life Science Discussion</description>
<language>en-us</language> 
<managingEditor>sci7feed@sci7.nojunkorherepleasespam.com</managingEditor>
<sy:updatePeriod>hourly</sy:updatePeriod>
<sy:updateFrequency>1</sy:updateFrequency>
<sy:updateBase>2005-05-05T12:00+00:00</sy:updateBase>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[Shut the computer and the instrument down, and start it back up. ...communication problems are common with 9000's.<br />]]></description>
<pubDate>Tue, 19 Jun 2012 19:01:58 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[ Hello I have a communications failure on the IEEE and turbo pump oil all over the place .more to follow....]]></description>
<pubDate>Thu, 14 Jun 2012 06:31:51 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[One thing I ran across not that long ago was an issue of the pulse stage and analog stage optimization settings not saving properly.  Basicaly I would run the pulse stage optimization, it would optimize and reset to a different voltage and it would appear to save, even when I switched over to a different workspace.  However as soon as I would start to analyze within the new workspace, it would revert back to the old settings.  It was really weird and easy to miss as well.  It...]]></description>
<pubDate>Tue, 18 Oct 2011 21:07:38 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[Are you getting &amp;quot;actual&amp;quot; counts or are they staying the same the whole way through the optimization? If you have counts and they are satying the same I would suggest looking at the capacitor. <br />Are you using argon for your pulse and analog? If you are and you sere counts that are changing normally, it is most likely a sample introduction problem. If the argon counts a not changing...look at your lens wired and your capacitor. <br />]]></description>
<pubDate>Tue, 11 Oct 2011 15:34:44 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[@ ToxLabRat<br />Thanks for your reply<br />We did not do any cleaning etc prior to this issue.. and we rechecked the wires to make sure eberything was intact. I am thinking about rechecking for loose wires just to make sure.<br />When we ran pulse and analog stage optimization, we are getting seeing counts in the realtime window. However, when we run a standard/dual detector solution, we are still not seeing any counts.<br />We also checked to make sure that there is flow through sample...]]></description>
<pubDate>Mon, 10 Oct 2011 06:48:43 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[Check your capacitor....it is in the little box (seperate screws and cover) behind your detector. I had the same problem (zero counts) and burn marks were visible on the board with the capacitor (it stemmed from a power problem). I know an electronics tech that can refurbish them if that is the problem.<br />]]></description>
<pubDate>Fri, 07 Oct 2011 12:08:17 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[@MSResTechTT:<br />Did you do anything at all to the instrument just prior to this issue occuring (I.e. clean lens, clean cones, etc)?  One common thing that causes you to have no counts is a failure to reconnect the lens power supply after cleaning or replacing your lens.  So if you did any work where you disconnected the lens, then I would go back and check that you did snap the wires back together.  I would also check and make sure that the wires are not coming loose from the...]]></description>
<pubDate>Thu, 06 Oct 2011 21:07:33 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[by slit, do you mean the gate valve?<br />if so then yes<br />there are no errors shown in any of the parameters when we enter service mode.]]></description>
<pubDate>Thu, 06 Oct 2011 07:09:50 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[Does your slit open after starting the plasma?<br />]]></description>
<pubDate>Tue, 04 Oct 2011 14:44:54 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[Hi all, I am experiencing a particular problem with our ELAN 9000. <br />Of recent, I am only obtaining zeros for all my signals. A few days before, I was able to light  the plasma, and run smart tune daily with passing, with no trouble whatsoever.<br />A couple days after, (after changing the argon tanks) I proceeded to do the same and the plasma lit with ease however in running smart tune daily, i noticed no values (all zeros) in the realtime window and of course, it failed.<br />We tried a...]]></description>
<pubDate>Tue, 04 Oct 2011 08:40:28 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[Thanks ToxLabRat for your reply!<br />Before reading your message, I noticed that the interface and the torch cower had some dust. I cleaned these parts with a clean cloth and then with Ar gas. After that, I checked the proper contact of the sensors' connectors (behind the torch box). Although the ELAN did not start properly at first, I reset the entire system (along with the PC) and it started normally. So far, there is no problem...<br />However, I will change the oil of the inte...]]></description>
<pubDate>Wed, 16 Mar 2011 13:24:29 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[The fault &amp;quot;RF Generator Communication Failure&amp;quot; is typically thrown when the oil in the interface roughing pump needs to be changed (The larger of the two mechanical pumps down at the bottom of the instrument).  So the first thing I would do is change your oil and then go ahead and re-set the system and see if that helps your issue.<br />As for the other issue with your main water flow state being low and especially with the pressure on your chiller being high, I would say t...]]></description>
<pubDate>Tue, 15 Mar 2011 19:15:14 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[Hello,<br />I'm coming here to see if any one can help me with a communication failure in my ELAN DRC-e from PE.<br />After a week with no use, I tried to start the ELAN, and I notted a Strange error alert that said: RF Generator communication failure &amp;gt;&amp;gt; RF generator OK, and it was repeated every minute. I re-started de PC and the Isntrument, and the problem goes out. Three days later, the problem appeared again, and remains untill now. <br />I followed the troubleshooting...]]></description>
<pubDate>Tue, 15 Mar 2011 13:28:32 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[It looks like you need to replace your big roughing pump Varian DS602.<br />But try new oil first.<br />Good Luck]]></description>
<pubDate>Sat, 16 Oct 2010 21:22:14 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[Communication failures where the system breaker is tripped (one of the four breakers located on the right hand side of the instrument at the end) are almost always related to the interface roughing pump (the larger of the two mechanical pumps under the instrument-the one on the right when facing the instrument).  When you looked at the oil, did you just assess the level of the oil or did you assess both the level and the color of the oil?  If you did not note the color, then ...]]></description>
<pubDate>Wed, 13 Oct 2010 00:42:26 GMT</pubDate>
</item>
<item>
<title>Elan 9000 ICPMS Communication Error</title>
<link>http://www.scientistsolutions.com/t16293-elan+9000+icpms+communication+error.html</link>
<description><![CDATA[Hi,<br />I'm having a little bit of difficulty with my ELAN 9000 system. We only use it twice a week (Tuesday + Thursday) and had no issues the last time I used it. Today when I tried and turn on the plasma, the roughing pump makes a loud grinding sound and immediately the system switch gets shut off, resulting in a communications failure.<br />I've rebooted the PC and the instrument. - No difference<br />Checked the oil - Oil looks fine. Was changed a month and a half ago on both vacuum pumps.<br />...]]></description>
<pubDate>Tue, 12 Oct 2010 14:39:49 GMT</pubDate>
</item>